Cookies auf den Omega Webseiten Wir verwenden Cookies auf unserer Webseite. Diese Cookies sind für die ordnungsgemäße Funktion der Webseite unerlässlich. Wenn Sie fortfahren, ohne Ihre Einstellungen zu ändern, gehen wir davon aus, dass Sie alle Cookies auf dieser Webseite akzeptieren.Um weitere Informationen zu diesen Cookies zu erhalten, klicken Sie bitte hier.
The OM-CP-THERMALERT-GB and OM-CP-THERMALERT-P are
temperature monitoring and alarming systems designed for
use in hospitals, laboratories, pharmaceutical facilities and other
critical temperature monitoring areas. Also, these units can be used
to monitor a single temperature location or expanded to include
hundreds of locations by adding additional transmitters. These units
record both ambient temperature and humidity.
The OM-CP-THERMALERT-GB and OM-CP-THERMALERT-P provides
real time notification of temperature deviations, and features a user
programmable alarm that can be configured to send a message via
text message, screen alarm and/or e-mail if an alarm condition occurs.
The OM-CP-THERMALERT-GB system includes an RTD probe that
is mounted in a miniature bottle filled with ethylene glycol. In this method,
a temperature “buffered” reading is taken, preventing false alarms.
The OM-CP-THERMALERT-P system includes an RTD probe with a 90 cm long lead wire and a 115 mm long probe sheath.
The OM-CP-THERMALERT-GB and OM-CP-THERMALERT-P system is
equipped with wireless two-way communication. Full communication
can be performed directly from a central PC, including starting,
stopping and downloading. In addition to wireless transmission of
data, both units store each reading to internal memory for backup. This
back up data may be retrieved wirelessly at a later time. Customers
can set-up automatic archiving preferences, ensuring all data is
saved and retained to comply with federal regulations.
Specifications
TEMPERATURE
Probe Temperature Range:
OM-CP-THERMALERT-GB: -50 to 80°C (-58 to 176°F) OM-CP-THERMALERT-P: -200 to 260°C (-328 to 500°F) Probe Calibrated Accuracy:
±0.1°C, -20 to 80°C; ±0.5°C (outside of that range) Resolution: 0.01°C (0.018°F)
WIRELESS
RF Frequency: 2.45 GHz IEEE
802.15.4 ultra-low power wireless
transceiver with fully bi-directional
communication Band: ISM band 2.405-2.48 GHz Maximum Output Power: +0dBm typical Receiver Sensitivity (OM-CP-RFC1000-EXT): -95dBm typical Range: Outdoor Line-of-Sight 610 metres Indoor: 152 metres
GENERAL
Reading Rate: One reading every
two seconds to one every 24 hours Memory: 30,000 readings; software
configurable memory wrap Wrap Around: Yes Start Modes: • Immediate start • Delay start up to 18 months Calibration: Digital calibration through software Calibration Date: Automatically recorded within device Battery Type: 3.6V lithium battery
(included); user replaceable Battery Life: 2 years typical Data Format: Date and time stamped °C, °F, K, °R Time Accuracy: ±1 minute/month (at 25°C) Computer Interface: OM-CP-RFC1000-EXT-EUR required; 115,200 baud Software: XP SP3/Vista/7 and 8 (32- and 64-bit) Operating Environment:
-20 to 80°C (-4 to 176°F), 0 to 95% RH non-condensing Dimensions:
Data Logger:
55 W x 127 L x 33 mm D Wire: 2.7 m Probe: 0.19 dia x 114 mm Enclosure Material: ABS plastic
*Alle Preisangaben in Euro. Anmerkung: Comes complete with 3.6V lithium battery.